Product Description
Analytical Flow Series cartridges: 350 µL and 500 µL are interchangeable and fit in all Analytical Flow Series In-line, Binary tee and Ternary tee housings..
Delay volume specified below was calculated based on an ASI engineering drawing. The delay volume can be verified by injecting a sample at a known flow rate, then measuring the time this sample first appears at the exit of the mixer.
Back pressure varies directly with flow rate. For example, if the back pressure is 2 PSI at 1 mL/min., it will be 4 PSI at 2 mL/min.
Specifications
Port size: 10-32 CPI (standard HPLC fittings for 1/16″ tube)
Compatible with Thermo Scientific™ nanoViper™ Fingertight Fittings
Maximum Operating Pressure: 6,000 PSI
Delay volume: 350 µL
Back pressure: < 2 PSI @ 1 mL/min. with water